Certified Tinted Samples

Nov. 20, 2013

Each meter is equipped with two test standards with light transmittance values printed on their labels, These values are accurate to within plus or minus one percentage point of a N.I.S.T. calibrated spectrophotometer.

To verify The Tint Meter is working properly and within specifications, take a reading of the reference samples before you begin to use the Tint Meter. The meter should display a reading of not more than plus or minus 3% points of the value printed on the label.

Request More Information

By clicking above, I acknowledge and agree to Endeavor Business Media’s Terms of Service and to Endeavor Business Media's use of my contact information to communicate with me about offerings by Endeavor, its brands, affiliates and/or third-party partners, consistent with Endeavor's Privacy Policy. In addition, I understand that my personal information will be shared with any sponsor(s) of the resource, so they can contact me directly about their products or services. Please refer to the privacy policies of such sponsor(s) for more details on how your information will be used by them. You may unsubscribe at any time.